OTG-MPK8 Fiber Optic Temperature Sensor

High accuracy SCBG-based fiber optic temperature sensor for tip measurement applications.

Opsens Solutions’ OTG-MPK8 fiber optic temperature sensor offers the highest performances and robust packaging in the industry. The OTG-MPK8 sensor uses the well proven technique based on the temperature-dependent bandgap of GaAs crystal as the temperature transduction mechanism. Its small sensing GaAs crystal located at the tip of the optical fiber makes it convenient for tip measurement applications. The OTG-MPK8 temperature sensor is housed in a robust PEEK tube hence facilitates insertion and daily handling.

Combined with Opsens Solutions’ SCBG (Semiconductor Bandgap) signal conditioning technology and with the inherent advantages of fiber optic, the OTG-MPK8 delivers unprecedented repeatability and reliability in the most adverse conditions such as high level of EM, RF, MR, NMR and microwave field environments.

The OTG-MPK8 is made with industry standard optical fiber and is compatible with all Opsens Solutions’ SCBG signal conditioners. This robust fiber optic temperature sensor is available with different optical cables and sheath options and is customizable according to customer specific applications or for OEM-type applications.

Benefits

  • ± 0,3 °C total system accuracy
  • Small and robust design
  • Outstanding repeatability
  • EMI/RFI and microwave immune
  • Intrinsically safe
  • OEM-type and custom disposable version

Applications

  • Surface and core temperature monitoring in EMI environments
  • RF and ultrasound environments
  • Temperature monitoring for preclinical and research applications
  • MR coil temperature monitoring
Temperature operating range

-20 °C to +250 °C (Higher temperature ranges available upon request)

Resolution

0.01 °C

Accuracy

± 0.3 °C (Total accuracy over the full range from +20 to 45 °C including both signal conditioner and sensor errors)
± 0.8 °C and better (Total accuracy on -20 to +250 °C including both signal conditioner and sensor errors)

Response time

500 ms or better

Operating humidity range

0-100 %

MRI/EMI/RFI susceptibility

Complete immunity

Calibration

NIST traceable

Cable length

1.5 meters standard (Other lengths available)

Optical connector

ST standard

Sensor diameter

0.810 mm O.D.

Cable sheathing

Teflon™ ( ≥ 85 °C)

Signal conditioner compatibility

All Opsens Solutions’ SCBG signal conditioners

Opsens Solutions Inc. reserves the right to make any changes to the above specifications without prior notice.